熱門關鍵詞:蘇瑪罐(guan),原子吸收分(fen)光(guang)光(guang)度計,環(huan)境監測(ce)產(chan)品,液相(xiang)色譜,色譜柱。
更新時間:2019-11-11
Dionex Atlas 電解抑制器為你的離子分離,尤(you)其是在使用(yong)時,碳酸洗脫液
免(mian)費咨詢(xun):010-61755520
為你的離子分離,尤其是(shi)在使(shi)用時,碳酸洗脫液,用的Thermo Scientific 戴(dai)安 AES Atlas電(dian)解(jie)抑(yi)制器更(geng)高的靈敏度和更(geng)低的噪音。你不會有混合再生劑;這抑(yi)制依賴于(yu)電(dian)解(jie)自動,連(lian)續再生。
Dionex Atlas 電解抑制器
訂貨信息:
Anion Atlas Electrolytic Suppressor | 056116 |
Cation Atlas Electrolytic Suppressor | 056118 |
The Thermo Scientific Dionex Atlas™ Electrolytic Suppressor (AES™) is a continuous electrolytically-regenerated suppressor based on the MonoDisc™ and AutoSuppression™ products. The unique design of the Dionex AES suppressor enables very fast startup times and an extremely low noise level, particularly for carbonate eluents. The benefits of the Dionex AES suppressor include:
High performance and very low noise for carbonate eluents
Low background conductivity, and excellent baseline stability
Easy to use maintenance-free operation
Fast startup and equilibration
Will suppress eluents up to 25 μeq/min
The MonoDisc suppression bed of the Dionex AES is composed of ion-exchange monolith and flow distribution disks. This configuration facilitates efficient exchange of the eluent counter ions for regenerant ions, resulting in eluent suppression and analyte response enhancement.
Atlas Electrolytic Suppressors Specifications | |
---|---|
AAES | |
Physical Specifications | |
Dimensions (l × w × d) | 4.9 × 4.4 × 10.2 cm (1.9 × 1.8 × 4.0 in.) |
Weight | 120 g (0.3 lb) |
Void volume | <35 μL |
Chemical Specifications | |
Temperature Range1 | 15–40° C |
Recommended Backpressure | 20–100 psi |
Maximum Eluent Flow Rate | 3 mL/min |
Eluent Solvent Restrictions2 | 0% solvents (no solvents) |
Maximum Regenerant EWM3 Flow Rate | 5 mL/min |
Modes of Operation Supported | Recycle and external water |
Maximum Current | 150 mA |
CAES | |
Physical Specifications | |
Dimensions (l × w × d) | 4.9 × 4.4 × 10.2 cm (1.9 × 1.8 × 4.0 in.) |
Weight | 120 g (0.3 lb) |
Void volume | <35 μL |
Chemical Specifications | |
Temperature Range1 | 15–40° C |
Recommended Backpressure | 20–100 psi |
Maximum Eluent Flow Rate | 3 mL/min |
Eluent Solvent Restrictions2 | 0% solvents (no solvents) |
Maximum Regenerant EWM3 Flow Rate | 5 mL/min |
Modes of Operation Supported | Recycle and external water |
Maximum Current | 150 mA |
1When installed outside the heated column enclosure, all suppressors can support applications up to 60 ?C.
2Solvents for anion eluents include methanol. Solvents for cation eluents include acetonitrile and dioxane.
3EWM = external water mode; for eluents containing > 40% solvent, use the chemical regeneration mode.